AirTrack-Mobile is a compact version of aerosol monitors in terms of advanced design solutions for automatic alpha/beta spectrometric measurements.
A special version of the Silicon Ion Implanted Detector with area 600 mm2 is used for alpha and beta spectrometry. The metalized coating of entrance window with thickness of less than 2 mm provides mechanical and chemical protection of the detector. Measurements can be carried out in high background conditions – compensation of radon, thoron and their progenies is performed algorithmically by spectrometric measurements. To reduce the influence of gamma background an auxiliary SIID is used to measure beta equivalent pulses and subtract them from the total beta counts.
A modular approach to the design of AirTrack-Mobile hardware and software enables fast troubleshooting of system issues. Moreover, modular designs allow easiness of upgrades and/or integration of future equipment. In general, the system and its components are ergonomically designed to ensure ease and comfort of operation.
Value | Parameter |
---|---|
Detector Unit | |
Detectors | 1) Silicon Ion Implanted Detector 600 mm2 2) Additional SIID for gamma background compensation |
Detection efficiency | 241Am-241 not less than 0.15 90Sr/90Y not less than 0.2 |
Alpha energy range | 2 MeV to 10 MeV |
Beta energy range | 80 keV to 3 MeV |
Alpha measuring range | from 10^-2 to 105 Bq/m3 |
Beta measuring range | from 10^-1 to 105 Bq/m3 |
Radon and Thoron measuring range | from 10^-1 to 105 Bq/m3 with automatic Radon progenies compensation |
MCA | 1024 channels |
Interface | |
Internal Data Storage | up to 240 GB SSD |
Display | LCD Touch Panels 5.7 (optional 6.5) |
Data transmission interfaces | USB, LAN, RS-485 |
Sensors | |
Air flow-rate | up to 6 m3/h (100 l/min) |
Differential pressure | up to 200 mbar through the filter |
Air inlet temperature | from -40 to +70 °C |
Pump unit | |
Nominal flow rate | 3 m3/h (50 l/min), adjustable manually |
Noise level (ISO 2151) | not more than 59 dB(A) |
Protection | IP54 |
Filter unit | |
Type | Filter tape |
Material | PTFE, (Glass Fiber, Cellulose - optionally) |
Tape Cassette | 15 meters lengths, 50 or 60mm wide, 90 days autonomy |
Automatic change | When total activities or pressure drop thresholds are exceeded |
Alarm Unit | |
Lamps | Red, Green LED with flash frequency 1 Hz |
Siren | 2 tone 85dB |
Protection category | IP66 |
Check sources | |
Am-241 & Sr-90 | Type EM X, activity from 2000 to 5000 Bq/h with certificate |
Operational characteristics | |
Voltage and frequency | 230 ± 20 V AC 50 Hz, 800 W max |
Temperature | from 0 °C to +40 °C, 85 % RH non-condensing |
Dimensions | 800 × 780 × 1575 mm |
Weight | no more than 60 kg |
Protection | IP 54 |