State of Health Preamplifier (SHP) for HPGe detectors

Application

State of Health Preamplifier (SHP) is a low noise and fast resistive-feedback preliminary amplifier intended for operation within gamma- and X-ray Detection Unit based on HPGe detector. SHP module allows to perform monitoring of detection unit basic parameters in real-time mode or per request of the operator. Specialized software SHP Center is used to display the parameters of the preamplifier in real time and control the settings. Analysis of the performance in the long run allows to predict possible failure of the detector and to perform timely preventive maintenance.

Features

SHP module provides information:
  • Preamplifier serial number
  • Detector serial number
  • Detector type
  • Cryostat model
  • Recommended HV value
  • HV polarity
  • Output signal polarity
  • Shaping time recommended value
  • Flat top time recommended value
  • ADC parameters
  • Recommended cooling time
  Measuring and display preamp parameters in real-time:
  • FET voltage
  • FET current
  • FET drain voltage
  • Preamplifier output voltage
  • Resistive-feedback output voltage of the preamplifier
  • Preamplifier power supply voltage
  • Detector leakage current
  • HPGe crystal temperature
  • SHP module internal temperature
  • SHP module internal humidity
  • SHP module internal atmospheric pressure
Control built-in test-pulser
  • Control test generator frequency
  • Control test generator amplitude
Software
  • SHP center
 

DESCRIPTION

State of Health Preamplifier (SHP) is a low noise and fast resistive-feedback preliminary amplifier intended for operation within gamma- and X-ray Detection Unit based on HPGe detector.

SHP module allows to perform monitoring of detection unit basic parameters in real-time mode or per request of the operator. Analysis of the performance in the long run allows to predict possible failure of the detector and to perform timely preventive maintenance.

SHP module is specially designed to operate together with BSI HPGe detectors via USB interface. Adjustment of SHP parameters is performed through the software package developed specially for preventive maintenance and service activities.

SPECIFICATION

ParameterValue
Energy rangefrom 3 keV to 10 MeV
Output signal polaritynegative
Output signal Rise time (µS)≤ 0.15
Output signal fall time (µS) ≥ 50
Conversion Gain mV/MeV220-250
Output Impedance (ohm)50
Max constant input count rate Co-602x 105 cps
Preamplifier power supply is ±12 V with 9 pin connector compatible with NIM standards
TTL signal to shut down the HV:
- detector warm -0V;
- detector cold: +5V
- HV INHIBIT – BNC
- Preamplifier and High voltage filter included
- Automatic high-voltage shutdown to protect preamplifier input FET.

2 LEDs on the separate cable end from preamplifier:
- GREEN – detector is cooled down
- RED – high count rate
Cable connectors:
- 9 pin – PA Power supply;
- 11 SHV - detector bias power supply;
- BNC - signal output of detector, UART-TTL
- USB.
Indicator warns of excessive counting rate

FOR ADDITIONAL INFORMATION CONTACT US